76
Repair Procedures
Welch Allyn Propaq Encore Vital Signs Monitor
5.
Press the side panel out of the rear chassis. There is an adhesive gasket around the
side panel which requires considerable force to separate.
Do not pry against
components
.
6.
Remove the tubing attached to the cuff hose connector. You can reuse this tubing
when installing the new side panel.
Installing the Side Panels
1.
Clean all gasket material from the rear chassis where the side panel is to be installed.
(The adhesive side of duct tape can be used to pull away old gasket residue.)
2.
Attach the tubing to the NIBP cuff hose connector.
3.
Route the bundle of cabling with the bead under the temperature connector housing
so that when the side panel is installed the connectors will extend out of the rear
chassis along the bottom of the chassis.
INSERT SMALL SCREWDRIVER
BETWEEN PLASTIC AND CLAMP
TO RELIEVE TENSION AGAINST
PLASTIC
ROUTE CABLE BUNDLE
REAR CHASSIS
ALONG BOTTOM OF
Содержание Propaq Encore
Страница 1: ...Propaq Encore Vital Signs Monitor Service Manual...
Страница 14: ...10 Overview Welch Allyn Propaq Encore Vital Signs Monitor...
Страница 50: ...46 Troubleshooting Welch Allyn Propaq Encore Vital Signs Monitor...
Страница 96: ...92 Repair Procedures Welch Allyn Propaq Encore Vital Signs Monitor...
Страница 111: ...Service Manual Technical Overview 107 Cabling Diagrams cabling diag...
Страница 120: ...116 Technical Overview Welch Allyn Propaq Encore Vital Signs Monitor...
Страница 126: ...122 Field Replaceable Units FRUs Welch Allyn Propaq Encore Vital Signs Monitor...
Страница 128: ...124 Manufacturable Test Equipment Welch Allyn Propaq Encore Vital Signs Monitor...
Страница 130: ...126 Dynatech Nevada Patient Simulator Modification Welch Allyn Propaq Encore Vital Signs Monitor...