Repair Procedures
Welch Allyn Propaq Encore Vital Signs Monitor
1.
Detach the two ribbon cables from the rear chassis:
•
Ribbon cable to P10 on the Recharger board, from the Digital board.
•
Ribbon cable to P11 on the Recharger board, from the Digital board.
2.
Separate the front chassis from the rear chassis (see figure below).
Note
The two halves are sealed by rubber O-ring gaskets that are recessed around the
perimeter of the rear chassis. If the gaskets become loose, replace them in the
perimeter before re-assembling the monitor.