30
Functional Verification
Welch Allyn Propaq Encore Vital Signs Monitor
12. Disconnect the CO
2
sensor from the monitor.
13. Turn off the monitor.
14. Log the test date and results in your department log.
Sidestream CO
2
Verification Procedure
The following procedures verify the operation of the sidestream CO
2
(SCO
2
) option.
1.
Allow the monitor to warm-up for 3 minutes before performing the SCO
2
tests.
2.
If present, disconnect the mainstream sensor from the monitor.
3.
If installed, remove the water tap from the receiver on the front left panel of the
expansion module
4.
Determine the local barometric pressure (
“CO2 Verification Preparation”
steps 1-3.
5.
Select the SIDE TST screen through SETUP > MORE > MORE > SERVICE > YES >
CO
2
TEST
WARNING
Eliminate all the possibilities causing errors shown in the following
table before performing the calibration procedure (
on page 42). Recalibrating the altimeter, as described in that
procedure, without adequately eliminating all other possible faults may result in
inaccurate EtCO
2
readings.
Table 9. Possible Sources of CO
2
Verification Test Errors
Error
Specification
Correction
Incorrect/unknown calibration
gas concentration
4 to 10% certified to within ±0.01% Verify gas supply. Gas must be between 4% and
10%. Once known, convert percentage to partial
pressure (mmHg) (
Insufficient/excessive gas
flow rate
Between 60 and 200 ml/min
Verify with flow meter.
Excessive back pressure in
cuvette
Back pressure should not exceed
1þmmHg
Connect gases as instructed in verification test.
Air mixed with standard gas
Check integrity of test setup.
Gas temperature
Gas temperature entering airway
adapter should be at room
temperature
Store gas at room temperature.
Cuvette error
Results should not vary more than
0.5 mmHg due to cuvette tolerances
Use another cuvette (airway adapter).
Thermal equilibrium error
Sensor not warmed up
Allow sensor/system to operate for a 15 minute
warm-up period.
High CO
2
reading with sensor
subject to room air only
CO
2
forced outside of airway adapter
into sensor. This is possible during
demonstration if exhaled gases are
blown through hands cupped around
cuvette and exhaled gases are
forced into sensor.
Contain exhaled breath through the inside of
cuvette only. Do not allow gases to be forced
outside of cuvette and into sensor housing. If
CO
2
has been forced into sensor allow 24 to 48
hours for the CO
2
to dissipate.
Содержание Propaq Encore
Страница 1: ...Propaq Encore Vital Signs Monitor Service Manual...
Страница 14: ...10 Overview Welch Allyn Propaq Encore Vital Signs Monitor...
Страница 50: ...46 Troubleshooting Welch Allyn Propaq Encore Vital Signs Monitor...
Страница 96: ...92 Repair Procedures Welch Allyn Propaq Encore Vital Signs Monitor...
Страница 111: ...Service Manual Technical Overview 107 Cabling Diagrams cabling diag...
Страница 120: ...116 Technical Overview Welch Allyn Propaq Encore Vital Signs Monitor...
Страница 126: ...122 Field Replaceable Units FRUs Welch Allyn Propaq Encore Vital Signs Monitor...
Страница 128: ...124 Manufacturable Test Equipment Welch Allyn Propaq Encore Vital Signs Monitor...
Страница 130: ...126 Dynatech Nevada Patient Simulator Modification Welch Allyn Propaq Encore Vital Signs Monitor...