100
Technical Overview
Welch Allyn Propaq Encore Vital Signs Monitor
Power Electronics
The power system circuits are located on the Analog, Digital, and Recharger boards. The
power system includes the ac mains power adapter.
The power system
•
converts ac mains current to dc current (external power adapter)
•
recharges the internal battery pack (Recharger board)
•
converts battery power to several voltages for the display (Recharger board)
•
converts battery power to several regulated and unregulated power supplies for
both isolated and non-isolated electronics (Analog board)
•
provides isolated power for RS423 communications (Digital board)
•
controls the activation and deactivation of the power system (Digital board)
Power is delivered from ac mains to the monitor through a power adapter. The adapter
converts ac to dc current at approximately 15 Volts.
The internal battery continually supplies monitor power and is recharged while the power
adapter is plugged into ac mains and connected to the monitor. The monitor can also be
recharged from any dc power source that meets the monitor’s dc power input
specifications. An accessory cable is available from Welch Allyn to allow connection of
such a power source to the dc power input connector on the monitor’s right side panel.
This cable must be used for monitor calibration.
Thermistors in the battery compartment monitor battery temperature. As battery charging
takes place, battery temperature rises. The recharger output voltage is temperature-
compensated; as battery temperature rises, maximum recharger output voltage
decreases.
Switching power supplies generate power for recharging the battery, for powering the
display, and for other power sources throughout the monitor. Suppression devices
designed into the power system provide EMI suppression. A transformer bridges the
battery power to the isolated electronics.
Several control lines activate and deactivate the power supplies. These control lines
originate on the Digital board and are routed to the Recharger board and Analog board to
activate the display power supply and the main power supply. The recharger supply is
automatically activated any time an adequate dc power source is connected to the power
input jack on the monitor’s right side panel.
Содержание Propaq Encore
Страница 1: ...Propaq Encore Vital Signs Monitor Service Manual...
Страница 14: ...10 Overview Welch Allyn Propaq Encore Vital Signs Monitor...
Страница 50: ...46 Troubleshooting Welch Allyn Propaq Encore Vital Signs Monitor...
Страница 96: ...92 Repair Procedures Welch Allyn Propaq Encore Vital Signs Monitor...
Страница 111: ...Service Manual Technical Overview 107 Cabling Diagrams cabling diag...
Страница 120: ...116 Technical Overview Welch Allyn Propaq Encore Vital Signs Monitor...
Страница 126: ...122 Field Replaceable Units FRUs Welch Allyn Propaq Encore Vital Signs Monitor...
Страница 128: ...124 Manufacturable Test Equipment Welch Allyn Propaq Encore Vital Signs Monitor...