Lead Fail Test
Perform the lead fail test as follows:
Connect the patient cable to a simulator or a shorting bar. Disconnect one lead at a time from the simulator or
shorting bar and verify that the correct lead fail message appears on the LCD while in the Main Menu screen. Only
the message for the disconnected lead should appear as shown below.
Lead
Message
Right Arm
LEAD RA OFF
Left Arm
LEAD LA OFF
Left Leg
LEAD LL OFF
V1
LEAD V1 OFF
V2
LEAD V2 OFF
V3
LEAD V3 OFF
V4
LEAD V4OFF
V5
LEAD V5 OFF
V6
LEAD V6 OFF
All Leads off
LEADS OFF
Transmitting Records
You may transmit ECGs from the ELI 250/210 to another Mortara Instrument electrocardiograph or to an E-Scribe
Data Management System using a direct connection or an optional factory installed internal modem.
Before transmitting ECGs to E-Scribe, the site number must be defined as well as the telephone number in the
configuration.
NOTE: Telephone transmission is available with modem option only.
To transmit records (either to another cardiograph or to E-Scribe), select the
XMT Key
. The following
screen is displayed:
To batch transmit of all records in the directory, select
F1 (Batch)
. If a batch transmission is selected, only those
records which have not been previously transmitted or marked deleted will be transmitted.
An ECG may be listed in the directory but have a “deleted status” (indicated by “X”). The directory saves those
records that have been marked for deletion in the event that you may want to recover the ECG at a later time. All
Содержание ELI 150
Страница 3: ......
Страница 10: ...viii...
Страница 14: ...xii...
Страница 36: ......
Страница 41: ...Section 4 4 5 Open the writer door by lifting up on the writer door latch and sliding the door to the left...
Страница 44: ...ELI 150 4 8 Re assemble in reverse order...
Страница 54: ...ELI 150 4 18 Re assemble in reverse order...
Страница 72: ......
Страница 74: ...ELI 150 5 2 ELI 150 Block Diagram...
Страница 77: ...Section 5 5 5 Specifications...
Страница 80: ......
Страница 100: ......