ELI 150
6-16
Set the “Default XMT Media” to WLAN and install your network information; the following settings are required
in the cardiograph and should be obtained from your local network personnel:
Host IP address
(network specific)
Port #:
5005
SSID:
(access point name)
Channel Number: 6
WEP info (if used)
Once this setup is completed ECGs can be transmitted by pressing the “XMT” key from the main menu and
selecting the record you wish to send.
Transmission information can be viewed in the test software window.
When completed the UUT display will return to the ECG directory and a “T” flag will be set indicating proper
transmission. The transmitted ECG can be seen in the target directory (C:\UNIPRO) listed as a new file.
Safety Tests
If the cardiograph housing was opened for repair or inspection work, the following safety tests should be performed in
accordance with the EN/IEC 60601-1 or EN 62353/IEC 62353 methods and limits.
In countries with different national legislation, the country specific standards and guidelines for retesting of electrical
medical devices are obligatory.
Objective evidence of testing results versus the defined standards should be provided as part of the servicing record.
The following tests are performed per the above specifications:
AC Hi-Pot Patient to Mains
Earth Leakage
Enclosure Leakage
Patient Source
Patient Sink
Содержание ELI 150
Страница 3: ......
Страница 10: ...viii...
Страница 14: ...xii...
Страница 36: ......
Страница 41: ...Section 4 4 5 Open the writer door by lifting up on the writer door latch and sliding the door to the left...
Страница 44: ...ELI 150 4 8 Re assemble in reverse order...
Страница 54: ...ELI 150 4 18 Re assemble in reverse order...
Страница 72: ......
Страница 74: ...ELI 150 5 2 ELI 150 Block Diagram...
Страница 77: ...Section 5 5 5 Specifications...
Страница 80: ......
Страница 100: ......