Section 3
3-9
Configuration Page 2
Flash Size
Flash Size indicates the ECG storage capacity of your electrocardiograph. Normal Flash Size is the standard
memory capacity. If the optional expanded memory has been installed, Expanded will be displayed.
ID Format
This option defines the format for the patient ID information prompts. There are three standard formats: short,
standard, or long. A custom ID format is optional when downloaded from the E-Scribe Data Management System.
The short format includes the patient's last name and first name, patient ID number, date of birth, which will
automatically calculate the age, and gender field.
The standard format includes the patient's last name, patient ID number, age, height, weight, gender, race,
medication 1, medication 2, and a location field.
The long format is identical to the standard format except that it includes the patient's first name, room and comment
fields.
The custom format, which is designed in the Mortara E-Scribe Data Management System, can be downloaded to the
ELI 150. The custom ID format is uniquely designed to meet your facility’s needs. Please see Appendix A to
download a Custom ID.
Auto-Fill ID
When enabled, the system will automatically populate the demographic fields in the ID screen. When the patient ID
field is manually populated and followed either by selecting enter or
F2 (▼)
the system automatically scans the
patient directory and if records with the exact ID are found, the existing data is used to fill some of the demographic
fields. The auto-fill feature is designed to only automatically populate last name, first name, date of birth, age and
gender. If no matching records are found, a brief message is displayed and the user must manually enter the
patient’s demographics.
NOTE: In order to avoid the use of incorrect data, the Auto-Fill ID feature is only supported if the ID
formats are the same between records.
AC Filter
Содержание ELI 150
Страница 3: ......
Страница 10: ...viii...
Страница 14: ...xii...
Страница 36: ......
Страница 41: ...Section 4 4 5 Open the writer door by lifting up on the writer door latch and sliding the door to the left...
Страница 44: ...ELI 150 4 8 Re assemble in reverse order...
Страница 54: ...ELI 150 4 18 Re assemble in reverse order...
Страница 72: ......
Страница 74: ...ELI 150 5 2 ELI 150 Block Diagram...
Страница 77: ...Section 5 5 5 Specifications...
Страница 80: ......
Страница 100: ......