10.2.2 Area around the product—not enclosed
NOTE21
If the pump is to be installed inside an enclosure, contact your Watson-Marlow representative for
advice.
The pump must always be accessible to facilitate additional installation, operation, maintenance, and cleaning.
Access points must not be obstructed or blocked.
Installation clearances are provided in the illustrations and explanation table below:.
Minimum
clearance
Comment
A
200mm
Install and remove the pumphead
B
User to define
User to define: The minimum clearance is based on the
l
Bend radius of users cables
l
Room to install and remove control cables for Relay Module optioned pumps
C
25mm
Additional clearance will be needed to Access the back of the pump for :
l
Information (serial number, product name)
l
Carry out an Earth bond test
Update the software using a USB connection
D
40mm
(PROFIBUS model
115mm)
The clearance is based upon a pump with a door at point D that can be opened or
closed to the front of the pump.
Additional clearance will be required for the :
l
Installation of control cables
l
Operate and view the screen and keypad.
10 INSTALLATION: PART 1 (PHYSICAL)
50