15.1.9.1 Change mode: Flow calibration
(Manual, PROFIBUS, Universal and Un only)
Recalibrate:
l
After changing pumphead
l
After changing process fluid
l
After changing any connecting pipework.
l
Periodically to maintain accuracy.
This pump displays flow rate in ml/min.
To calibrate pump flow:
Procedure
1. Highlight
Flow Calibration
2.
SELECT
.
15 INSTALLATION: PART 6 (SETTING UP THE PUMP (CONTROL SETTINGS))
131