11.3.5 Pre-electrical installation checklist
Carry out the following pre-electrical installation check. At this point in the full installation procedure, the fluid
path or pumphead should not yet be installed.
l
Ensure then pump has been physically installed in accordance with
"10 Installation: Part 1 (Physical)" on
l
Ensure the power cable is not damaged
l
Ensure the electrical isolation device is installed, tested and ready for operation.
l
Ensure overcurrent protection is installed, tested and ready for operation.
If there is a problem with any of the above do not carry on with the electrical installation and instruct that the
pump is removed from service until the until the pre-electrical installation requirements have been met.
11.3.6 Connecting to a DC power supply
1. Carry out the pre installation checks in the previous procedure
2. Connect to DC power supply via the pre-fitted M8 Ring terminals.
l
Connect the red wire to positive (+)
l
Connect the black wire to negative (-)
If the pump is connected in reverse (reverse polarity), the pump will not power up. It will not create a hazard,
correct connection polarity and continue.
11.4 Testing of electrical power and first time pump
start up
11.4.1 Model: Remote
When electrical power is supplied to pump, all LED icons will illuminate for three seconds.
11.4.2 Model: Manual, PROFIBUS, Universal, Un
When the pump is turned on for the very first time a leak detection message will appear. This is because the
pumphead has not yet been installed. For the purposes of testing the electrical power to the pump, this message
indicates the pump is receiving power. The procedure for installing the pumphead for the first time is provided in
the next section.
11 INSTALLATION: PART 2 (ELECTRICAL POWER)
61