17.3.3.1 Removing the pumphead
Procedure
1. Stop the pump.
2. Isolate the pump from electrical power.
3. Drain down the fluid path in accordance with your organisation’s procedure
4. Remove the
inlet
and
discharge
fluid path connections pumphead (protecting the pump from any
process fluid spillages), by unscrewing the connection collars and gently pull the connections off the
pumphead ports. See illustration below.
Procedure
5. Release the pumphead locking lever.
17 MAINTENANCE
187