Procedure
4. Choose calibration method:
l
MANUAL
method—Enter value using
+/-
keys.
Or
l
INPUT
method—Apply current signals electrically to analog input.
Setting a high signal
Procedure
5.
MANUAL
—Enter value using
+/-
keys
INPUT
—Send high signal input to pump.
15 INSTALLATION: PART 6 (SETTING UP THE PUMP (CONTROL SETTINGS))
138
Содержание qdos 20
Страница 40: ...6 Storage 6 1 Storage conditions 41 6 2 Shelf life 41 6 STORAGE 40...
Страница 169: ...Procedure 9 NEXT 10 Use keys to set Alert Level Procedure 11 SELECT to return to FLUID LEVEL SETTINGS 16 OPERATION 169...
Страница 192: ...NOTICE The pumphead locking lever is designed to be loosened or tightened by hand 17 MAINTENANCE 192...