Figure 3–13: Probe assembly fitted to the probe adaptor
Probe adaptor fitting
Tip:
The probe adaptor fitting on the UniSpray probe assembly is not compatible with the
probe adaptor. If you cannot fit the probe adaptor cap, ensure that you are installing the
correct probe assembly.
3.
Tilt the probe adaptor cap, so that the ball-bearing is located in the recess at the bottom of
the aperture, and insert the probe assembly tubing through the aperture.
Figure 3–14: Probe adaptor cap
Probe cap aperture from the underside
Probe cap aperture from the top
Tip:
The probe assembly tubing can pass through the aperture only when the ball-bearing
is located in the recess at the bottom and does not block the aperture.
June 9, 2016, 715004990 Rev. A
Page 56