4.
Rotate and lift the tool and collar to remove the sample cone from the cone gas nozzle.
Figure 4–11: Removing the sample cone
Rotate the tool
Remove the sample cone
5.
Remove the O-ring from the sample cone.
Figure 4–12: O-ring removed from the sample cone
Sample cone
O-ring
Cone gas nozzle handle
Cone gas nozzle
June 9, 2016, 715004990 Rev. A
Page 107