Second ion guide assembly
Brown PEEK gasket
4.
Remove the O-ring from the differential pumping aperture on the second ion guide
assembly (see
).
Figure 4–30: StepWave second ion guide assembly
Differential pumping aperture
O-ring
Warning:
To avoid spreading contamination, dispose of all waste materials according to
local environmental regulations.
5.
If the O-ring shows signs of deterioration or damage, dispose of it in accordance with local
environmental regulations.
Notice:
To avoid misalignment of the StepWave ion guide, do not remove the
differential pumping aperture or any further components from the assembly.