• Source enclosure seal
• Nebulizer gas seal
• Desolvation gas seal
4.22.1 Removing the probe adjuster assembly probe and source
enclosure seals
Required materials
• Chemical-resistant, powder-free gloves
• O-ring removal kit
To remove the probe adjuster assembly probe and source enclosure seals:
Warning:
To avoid personal contamination with biohazards, toxic materials, and corrosive
materials, wear chemical-resistant gloves when performing this procedure.
1.
Remove the source enclosure from the instrument (see
).
2.
Use the O-ring removal kit to carefully remove the following seals from the probe adjuster
assembly:
• Probe seal
• Nebulizer gas seal
See also:
Figure 4–49: Probe adjuster assembly seals