6.
Using both hands, fit the source enclosure to the two supporting studs on the source
adaptor housing.
7.
Close the source enclosure.
Rationale:
Fitting and closing the source enclosure prevents debris entering the
instrument while you are working on the StepWave ion guide assembly.
4.15.4 Disassembling the StepWave ion guide assembly
Warning:
To avoid personal contamination with biohazards, toxic materials, and corrosive
materials, wear chemical-resistant gloves when performing this procedure.
Required materials
• Chemical-resistant, powder-free gloves
• Combined 2.5-mm hex wrench and cone extraction tool
• O-ring removal kit
To disassemble the StepWave ion guide assembly:
1.
Use the combined 2.5-mm hex wrench and cone extraction tool to remove the two screws
securing the first ion guide assembly to the second ion guide assembly.
Figure 4–27: StepWave ion guide assembly
June 9, 2016, 715004990 Rev. A