Figure 4–45:
Ion block cover plate securing screw
Ion block cover plate
4.
Remove the ion block cover plate.
5.
Use the combined 2.5-mm Allen wrench and cone extraction tool to loosen the captive
PEEK terminal block securing screw.
Figure 4–46:
PEEK terminal block securing screw
Heater cartridge assembly wires
Notice:
To avoid damaging the heater cartridge assembly wires, do not bend or
twist them when removing the assembly and ceramic heater mounting block from
the ion block.