Figure 4–7:
Ion block assembly
Notice:
To avoid damaging the StepWave assembly, do not switch the
instrument into Operate mode when the isolation valve is closed.
Notice:
To avoid damage, do not open the source isolation valve before fitting
the sampling cone assembly to the ion block assembly
4.13.2 Disassembling the sampling cone assembly
Required materials
• Chemical-resistant, powder-free gloves
• Combined 2.5-mm hex wrench and cone extraction tool
To disassemble the sampling cone assembly:
Warning:
To avoid personal contamination with biohazards, toxic materials, and corrosive
materials, wear chemical-resistant gloves when performing this procedure.
1.
Retrieve the combined 2.5-mm hex wrench and cone extraction tool from its storage
location behind the source enclosure.