Figure 4–31: Cleaning the first ion guide PCB assembly
First ion guide PCB assembly
Hook
3.
Add Waters MS Cleaning Solution to the glass vessel until the first ion guide PCB
assembly is immersed completely.
4.
Repeat step 1 through step 3 for the second ion guide PCB assembly, placing the hook
through one of the support rod holes.
Figure 4–32: Cleaning the second ion guide PCB assembly
Hook
Second ion guide PCB assembly
5.
Place the vessels containing the first ion guide and second ion guide PCB assemblies in
the ultrasonic bath for 20 minutes.
June 9, 2016, 715004990 Rev. A
Page 131