First-use tightening:
2.3.1.2 Stainless steel (gold-plated) fitting with long flats and two-piece stainless steel
ferrule (V-detail)
First use
Long flats
Compression screw
Two-piece stainless steel ferrule
Tighten the fitting finger-tight plus an additional 3/4-turn using a 1/4-inch open-end wrench.
Tip:
To prevent band spreading, ensure that the tubing is fully bottomed in the connection port
before you tighten the compression screw.
First use tightening
December 4, 2019, 715005052 Version 02 (previously released as Rev.B)