VTI Instruments Corp.
126
EX1200A Web Page Operation
Range Delay
: The amount of time the DMM takes between measurements on the same
function, but on different ranges.
o
Auto
checkbox: When selected, the auto range function is enabled.
o
Delay (sec)
: Sets the range delay. Valid values are between 0 and 4294.967295 s.
Auto Zero
: When enabled, the DMM automatically tares itself against an internal reference
before each reading.
Power Line (Hz)
drop menu: Selects either
50.0
Hz or
60.0
Hz as the ac frequency of the
DMM’s power source.
Aperture Time
: Indicates the time, in seconds, over which the DMM will integrate the
analog inputs. The longer the time, the more accurate the reading. Note that this property and
Digits
are interrelated hardware settings. Thus, changing one affects the other. Valid values
vary depending on the power line frequency and the mode setting.
o
For
60
Hz PLF:
0.00167
,
0.0167
,
0.167
, and
1.67
.
o
For
50
Hz PLF:
0.0002
,
0.002
,
0.02
,
0.2
, and
2.0
.
o
For
Frequency
and
Period
modes:
0.0014
,
0.014
, and
0.14
.
Saved Configurations Field
The
Saved Configurations
field allows the user to store, recall, and remove predefined settings
for the DMM. To save a configuration, simply define the configuration using the DMM soft from
panel, then click on the
Save
button in the
Saved Configurations
field. A prompt will appear
requiring a name for the configuration. Once a name is entered (only alphanumeric characters,
dash (-), and underscore(_) are allowed), click the
OK
button in the prompt to save these settings.
To recall a saved configuration, press the
Load
button, select the name of the configuration that is
required, then click on the
OK
button. To delete a preconfigured state, simply click the
Delete
button, select the state, the press
OK
. All saved configurations can be easily cleared at once by
selecting “Delete All Stored Configurations” from the list in the
Delete
dialog.
Measurement Field
From the
Measurement
field, the user can start a measurement.
Initiate
: This button initiates a measurement, moving the internal
state machine from ‘Idle’ to
‘Waiting for Trigger’.
Software Trigger Button
The
Send
button will send a software trigger to the DMM. This causes the trigger model to
advance if the selected
Sample Trigger Source
or
Trigger Input Source
is
Software
.
Readings Field
Below the DMM configuration fields, a data field is provided, with options, that is populated as
measurements taken by the EX1200A
’s DMM
are read from FIFO.
No.
: Indicates the row number, for reference.
Time
: Indicates the IEEE 1588 time the measurement was taken.
Value
: Displays the measurement data.
Unit
: Indicates the engineering units of the measured data.
Function
: Indicates the function used when taking the measurement was taken.
Option
: Indicates the function options (if applicable) used when the measurement was taken.
Range
: Indicates the range that was used when the measurement was taken.
Limits
: Indicates the trigger output limits met when the measurement was taken.
Содержание Ametek EX1200A Series
Страница 17: ...VTI Instruments Corp 16 EX1200A Series Introduction EX1200A SERIES 1 U FULL RACK MAINFRAME DIMENSIONAL DIAGRAM ...
Страница 18: ...www vtiinstruments com EX1200A Series Introduction 17 EX1200A SERIES 3 U MAINFRAME DIMENSIONAL DIAGRAM ...
Страница 33: ......
Страница 51: ...VTI Instruments Corp 50 EX1200A Series Preparation for Use ...
Страница 64: ...www vtiinstruments com EX1200A Series Preparation for Use 63 ...
Страница 65: ......
Страница 73: ......
Страница 107: ...VTI Instruments Corp 106 EX1200A Triggering ...
Страница 109: ...VTI Instruments Corp 108 EX1200A Triggering FIGURE 6 1 EX1200A SERIES TRIGGER MODEL ...
Страница 143: ...VTI Instruments Corp 142 EX1200A Web Page Operation ...
Страница 145: ...VTI Instruments Corp 144 EX1200A Series Programming ...
Страница 155: ......
Страница 161: ......