VTI Instruments Corp.
28
EX1200A Series Introduction
E
XPLANATION OF
S
PECIFICATIONS
This section provides explanatory notes for EX1200A DMM specifications that may be
misunderstood.
Memory
This is the internal FIFO/buffer memory of the EX1200A mainframe. This does not include any
buffer memory that the individual plug-in cards may have. Some plug-in cards, such as the
EX1200-7500, EX1200-3608, and EX1200-7416, have onboard buffers that allow them to
perform high-speed operations.
Saved States
The DMM can store its configurations inside a non-volatile memory. Recalling these
“saved
states” directly
from the memory saves programming effort and will speeds up the scanning
process.
Measurement Integration Time
Unlike data acquisition instruments that measure instantaneous values, the EX1200A DMM uses
an integrating ADC for measurements which allows it to take multiple samples and average them
by integrating them over a fixed period of time. This time is referred to as the
aperture time
and is
usually measured in power line cycles (PLC) counts. One PLC is equal to the time period of the
mains power voltage waveform (i.e., for a 60 Hz mains power, 1 PLC = 16.67 ms, whereas for a
50 Hz mains power, it is 20 ms). If the integration time is set to a multiple of PLCs, these errors
(and their harmonics) will average out to approximately zero. This is, however, only applicable for
dc measurements.
True RMS
The internal mainframe DMMs measure the root mean square (RMS) value of the applied
waveform unlike the “
a
verage measurement” instr
uments. As a result, the values returned by the
EX1200A DMM are closer to the readings of an analog instrument. A complex algorithm
calculates the true RMS value of the applied waveforms (refer to the
errors related to measuring non-sinusoidal waveforms). The inputs are always ac coupled, thus
removing all dc components.
AC Signal Filters
The EX1200A DMM has three different ac filters to allow for a balance between low frequency
accuracy and faster settling times for ac voltage and current measurements. Selection of the
appropriate filter improves the accuracy of reading as well as the scanning speed. Remember that
the minimum frequency input of the Low filter is 3 Hz. This means that the dc component of non-
symmetrical waveform inputs (such as pulse trains) are rejected before true RMS measurements
are made.
Power Supply
Each EX1200A mainframe has an inherent limitation on the maximum power it can supply to the
plug-in cards/modules. Please refer to the power consumption specifications of the plug-in cards
before configuring the cards. This data can be found in the
EX1200A
Switch Cards User’s
Manual
. Exceeding these power limits may damage the power supply of mainframe.
Содержание Ametek EX1200A Series
Страница 17: ...VTI Instruments Corp 16 EX1200A Series Introduction EX1200A SERIES 1 U FULL RACK MAINFRAME DIMENSIONAL DIAGRAM ...
Страница 18: ...www vtiinstruments com EX1200A Series Introduction 17 EX1200A SERIES 3 U MAINFRAME DIMENSIONAL DIAGRAM ...
Страница 33: ......
Страница 51: ...VTI Instruments Corp 50 EX1200A Series Preparation for Use ...
Страница 64: ...www vtiinstruments com EX1200A Series Preparation for Use 63 ...
Страница 65: ......
Страница 73: ......
Страница 107: ...VTI Instruments Corp 106 EX1200A Triggering ...
Страница 109: ...VTI Instruments Corp 108 EX1200A Triggering FIGURE 6 1 EX1200A SERIES TRIGGER MODEL ...
Страница 143: ...VTI Instruments Corp 142 EX1200A Web Page Operation ...
Страница 145: ...VTI Instruments Corp 144 EX1200A Series Programming ...
Страница 155: ......
Страница 161: ......