VTI Instruments Corp.
88
EX1200A Triggering
Noise Caused by Ground Loops
When measuring voltages in circuits where the DMM and the DUT are referenced to a common
earth ground,
but at different points, a “ground loop” is formed. As shown
in Figure 4-8, any
voltage difference between the two ground reference points (V
ground
) causes a current to flow
through the measurement leads. This causes errors such as noise and offset voltage (usually power
line related) which are added to the measured voltage.
R
L
= DUT Source Resistance Imbalance
R
i
= Multimeter Isolation Resistance
V
ground
= Voltage Drop on Ground Bus
V
TEST
R
L
HI
LO
DMM
R
i
> 10 G
V
ground
R
L
F
IGURE
4-8:
A
G
ROUND
L
OOP
The best way to eliminate ground loops is to maintain the
DMM’
s input isolation from earth
ground by not connecting the input terminals to ground. If the DMM must be earth-referenced, be
sure that the DMM and the DUT are connected to the same common ground point. This will
reduce or eliminate any voltage difference between the devices. It is also recommended that the
DMM and DUT be connected to the same electrical outlet whenever possible.
AC
V
OLTAGE
M
EASUREMENTS
The EX1200A DMM can measure the true RMS ac voltages within the limits specified. AC
voltage (ACV) measurements differ from the DCV measurements because simple integration and
averaging of the input signal will not yield the RMS value.
Taking an ACV measurement uses the same set up as a DCV measurement, with the HI and LO
leads being connected to the source. Be certain to observe the maximum specification ratings
before connecting to the DMM. As with DCV measurements, the source can be connected through
the front panel or through the DMM Bus backplane bus via a switch multiplexer module or the
DMM bus rear panel connector.
DMM Front Panel
Sense HI
Input HI
Input LO
Sense LO
AC Voltage Source
F
IGURE
4-9:
DMM
C
ONNECTIONS FOR
ACV
M
EASUREMENTS
Содержание Ametek EX1200A Series
Страница 17: ...VTI Instruments Corp 16 EX1200A Series Introduction EX1200A SERIES 1 U FULL RACK MAINFRAME DIMENSIONAL DIAGRAM ...
Страница 18: ...www vtiinstruments com EX1200A Series Introduction 17 EX1200A SERIES 3 U MAINFRAME DIMENSIONAL DIAGRAM ...
Страница 33: ......
Страница 51: ...VTI Instruments Corp 50 EX1200A Series Preparation for Use ...
Страница 64: ...www vtiinstruments com EX1200A Series Preparation for Use 63 ...
Страница 65: ......
Страница 73: ......
Страница 107: ...VTI Instruments Corp 106 EX1200A Triggering ...
Страница 109: ...VTI Instruments Corp 108 EX1200A Triggering FIGURE 6 1 EX1200A SERIES TRIGGER MODEL ...
Страница 143: ...VTI Instruments Corp 142 EX1200A Web Page Operation ...
Страница 145: ...VTI Instruments Corp 144 EX1200A Series Programming ...
Страница 155: ......
Страница 161: ......