The following control options are provided:
Get Continuous
checkbox: If enabled, this button disables the
Get
button and continually
populates the data table as data is acquired.
Get
button: When clicked, data is immediately pulled from FIFO memory.
Get All
checkbox: When enabled, clicking the
Get
button will retrieve all of the current data.
Once the FIFO is empty, it will stop acquiring retrieving data.
Get Count
field: Indicates the number of data points that will be returned when the
Get
button is clicked. Should be less than or equal to the
FIFO Count
.
FIFO Count
field: Indicates the number of data points available in the FIFO memory.
Clear Table
button: When clicked, all of the data currently in the table is erased.
Clear FIFO
button: When clicked, any data points currently stored in the FIFO memory will
be erased.
Save Data
button: When clicked, the data in the table can be saved. A .csv file is generated by
default.
DMM
C
ALIBRATION
W
EB
P
AGE
When the
Calibration
tab is clicked on the DMM SFP, a screen appears that allows the user to
access the calibration and verification routines used during production test.
Calibration Functions Field
The
Calibration Functions
field defines the type of calibration which will be performed as well
as they functions which will be calibrated. When
Calibration
is selected, new calibration
constants (adjustments) are generated and saved. The
Verify 24-hour spec
and
Verify 1-year
spec
options verifies the DMM’s
6½ digit specifications against its published 24-hour or 1-year
values, respectively.
Simple Verification
is similar to
24-hour
and
1-year
verifications, but uses
looser specifications than those published
. It is useful as a quick “sanity test” to ensure the
hardware is working properly.
Calibration Functions
drop menu: Select between the
Calibration
,
Simple Verification,
Verify 24-hour spec
, and
Verify 1-year spec
from this drop menu.
Содержание Ametek EX1200A Series
Страница 17: ...VTI Instruments Corp 16 EX1200A Series Introduction EX1200A SERIES 1 U FULL RACK MAINFRAME DIMENSIONAL DIAGRAM ...
Страница 18: ...www vtiinstruments com EX1200A Series Introduction 17 EX1200A SERIES 3 U MAINFRAME DIMENSIONAL DIAGRAM ...
Страница 33: ......
Страница 51: ...VTI Instruments Corp 50 EX1200A Series Preparation for Use ...
Страница 64: ...www vtiinstruments com EX1200A Series Preparation for Use 63 ...
Страница 65: ......
Страница 73: ......
Страница 107: ...VTI Instruments Corp 106 EX1200A Triggering ...
Страница 109: ...VTI Instruments Corp 108 EX1200A Triggering FIGURE 6 1 EX1200A SERIES TRIGGER MODEL ...
Страница 143: ...VTI Instruments Corp 142 EX1200A Web Page Operation ...
Страница 145: ...VTI Instruments Corp 144 EX1200A Series Programming ...
Страница 155: ......
Страница 161: ......