Trigger Delay
: indicates the delay, in seconds, that occurs between connecting a switch
channel to the DMM and when the measurement is taken. Valid values are
0
s to
4294.967295
s. An option to select
Auto
appears when this field is selected. Auto delay
allows the DMM to automatically set the trigger delay based on the measurement
configuration.
Gain
: Indicates the
m
in an
m
x+
b
calculation performed on the measurement data before it is
returned.
Offset
: Indicates the
b
in an
m
x+
b
calculation performed on the measurement data before it is
returned.
New
button: Adds a new DMM configuration.
Delete
button: Removes the selected DMM configuration.
Done
button: Closes the DMM Configuration Manager dialog box.
Notes on Web Page Use
The
DMM Configuration Manager
only shows the most common options. To fully configure the
DMMs, select the
DMM
tab and use its
Save
/
Load
/
Delete
buttons.
Trigger Properties Field
From the
Trigger Properties
field, the user configures the various triggering capabilities of the
scan engine.
Scan/Step Trigger In
: For the Scan Trigger, the scan engine waits for the selected event to
occur before starting each iteration of the scanlist. For the Step Trigger, the scan engine will
wait for the selected event to occur before starting each scan-step.
o
Immediate
: When selected, the scanner proceeds through immediately, without waiting
for any external signal.
o
Software
: When selected, the scanner proceeds through the scanlist in steps and will wait
for a software trigger before executing the next scan step or scanlist iteration.
o
LXI0
–
LXI7
: Indicates the LXI Trigger bus line that will advance the scanner to the
next step or scanlist iteration.
o
LAN0
–
LAN7
: Indicates the LAN event that will advance the scanner to the next step or
scanlist iteration.
o
DIO0
–
DIO7
: Indicates the digital I/O channel that will advance the scanner to the next
step or scanlist iteration.
o
Alarm0
: Indicates the Alarm event that will advance the scanner to the next step or
scanlist iteration.
Notes on Web Page Use
It is important to only send triggers other than Immediate when the EX1200A is expecting to
receive triggers. If the period of an alarm is set too fast, if the
Send Software Trigger
button is
clicked when the EX1200A is not in the
Waiting for Trigger
state, etc. the scan engine will flag
an error and abort the scan.
To wait for a trigger between measurements, place these triggers in separate scan steps (i.e.,
separate rows in the
Scanlist
table). The
Step Trigger In
field triggers advances between scan
steps.
Содержание Ametek EX1200A Series
Страница 17: ...VTI Instruments Corp 16 EX1200A Series Introduction EX1200A SERIES 1 U FULL RACK MAINFRAME DIMENSIONAL DIAGRAM ...
Страница 18: ...www vtiinstruments com EX1200A Series Introduction 17 EX1200A SERIES 3 U MAINFRAME DIMENSIONAL DIAGRAM ...
Страница 33: ......
Страница 51: ...VTI Instruments Corp 50 EX1200A Series Preparation for Use ...
Страница 64: ...www vtiinstruments com EX1200A Series Preparation for Use 63 ...
Страница 65: ......
Страница 73: ......
Страница 107: ...VTI Instruments Corp 106 EX1200A Triggering ...
Страница 109: ...VTI Instruments Corp 108 EX1200A Triggering FIGURE 6 1 EX1200A SERIES TRIGGER MODEL ...
Страница 143: ...VTI Instruments Corp 142 EX1200A Web Page Operation ...
Страница 145: ...VTI Instruments Corp 144 EX1200A Series Programming ...
Страница 155: ......
Страница 161: ......