EX1200A Triggering
95
The resolution of frequency measurements is set from one of the following options:
Slow Mode
,
Medium Mode
,
Fast Mode
, and
User Mode
. Table shows the settings used for each mode.
Mode
Digits
Aperture/Gate Time
Slow
6½
140 ms
Medium
5½
14 ms
Fast
4½
1.4 ms
User
4½, 5½, or 6½
T
ABLE
4-6:
F
REQUENCY
R
ESOLUTION
M
ODES
NOTE
Larger aperture/gate times increase the number of input signal cycles averaged for a reading. This
improves reading accuracy, but reduces scan speed. The DMM keeps the gate open for one full
waveform cycle up to the timeout value. So, if the input frequency is 3 Hz, the time needed for the
measurement reading to be returned can be up to 0.67 s. The default timeout is 1 s.
When in
User
mode, the following limits must be observed when setting the aperture time:
User Mode
Resolution (Hz)
Aperture Time (s)
Min
0.0025
16.67 x 10
-6
s
Max
42e-9
1 s
T
ABLE
4-7:
R
ESOLUTION AND
A
PERTURE
L
IMITS FOR
U
SER
M
ODE
Frequency and period measurement rates are dictated by the aperture time. The settling time
(trigger delay) is the same as the ACV slow mode setting. Note that the maximum measurement
time is no more than the following:
T
2
Time
Aperture
Time
t
Measuremen
where
T
= period of input waveform
Also note from Table 4-6 that aperture and resolution have an inverse relationship and can be
interchanged using the following formula:
(Hz)
Resolution
1
10
24
(seconds)
Time
Aperture
6
Connecting to the DMM
For front panel measurements, the ac signal should be connected to the HI and LO input jacks of
the EX1200A DMM using suitable connectors. Although not preferred in fixed installation,
measurements of this type can be used to quickly make readings and troubleshoot field wiring
issues.
DMM Front Panel
Sense HI
Input HI
Input LO
Sense LO
Frequency Source
T
ABLE
4-8:
DMM
C
ONNECTIONS FOR
F
REQUENCY
M
EASUREMENTS
Содержание Ametek EX1200A Series
Страница 17: ...VTI Instruments Corp 16 EX1200A Series Introduction EX1200A SERIES 1 U FULL RACK MAINFRAME DIMENSIONAL DIAGRAM ...
Страница 18: ...www vtiinstruments com EX1200A Series Introduction 17 EX1200A SERIES 3 U MAINFRAME DIMENSIONAL DIAGRAM ...
Страница 33: ......
Страница 51: ...VTI Instruments Corp 50 EX1200A Series Preparation for Use ...
Страница 64: ...www vtiinstruments com EX1200A Series Preparation for Use 63 ...
Страница 65: ......
Страница 73: ......
Страница 107: ...VTI Instruments Corp 106 EX1200A Triggering ...
Страница 109: ...VTI Instruments Corp 108 EX1200A Triggering FIGURE 6 1 EX1200A SERIES TRIGGER MODEL ...
Страница 143: ...VTI Instruments Corp 142 EX1200A Web Page Operation ...
Страница 145: ...VTI Instruments Corp 144 EX1200A Series Programming ...
Страница 155: ......
Страница 161: ......