www.vtiinstruments.com
EX1200A Web Page Operation
129
Temperature Field
The
Temperature
field provides the internal and external temperature during calibration.
Internal
: Displays the current internal temperature of the DMM.
External
: The current ambient temperature should be entered here when a calibration or
verification procedure is started.
Calibration Status Field
The
Calibration Status
field provides an indication of how close calibration is to completion.
Begin calibration by clicking on the
Initiate
button in this field.
Results Panel
The
Results
panel provides access to the results of the calibration or verification operation.
Calibration/verification begins by clicking on the
Initiate
button. The following options are
provided in the panel:
Technician
: This field must be filled in prior to saving the results data. It identifies the person
who performed this calibration or verification.
Reason
: This field must be filled in prior to saving the results data. The contents should
explain the reason the operation was performed, for instance, a reference to a support case or
RMA number.
Calibration Results
: When a calibration is performed, the
Results
field is populated as the
calibration procedure progresses and the following information is provided.
o
Function
: Indicates the DMM function to which the adjustment applies.
o
Option
: Indicates the option of the DMM function to which the adjustment applies.
o
Item
: Indicates the adjustment type that was made (gain, offset, etc.).
o
Range
: Indicates the measurement range to which the adjustment applies.
o
Minimum
: Indicates the smallest allowed value of the adjustment.
o
Maximum
: Indicates the largest allowed value of the adjustment.
o
Value
: Indicates the calculated value of the adjustment.
o
Pass/Fail
: If the calculated adjustment is within tolerances,
Pass
is indicated. If the
adjustments are not within tolerances,
Fail
is indicated.
Verification Results
: When a verification is performed, the
Results
field is populated as the
verification procedure progresses and the following information is provided.
o
Function
: Indicates the DMM function being tested.
o
Option
: Indicates the DMM function option being tested.
o
Range
: Indicates the DMM range being tested.
o
Magnitude
: Indicates the ideal magnitude for the given test point.
o
Frequency
: Indicates the ideal frequency for the given test point.
o
Reading
: Indicates the value measured by the DUT.
o
Reference
: Indicates the value measured by the reference DMM or reported by the
calibrator.
o
Minimum
: Indicates the smallest allowed value of the reading.
o
Maximum
: Indicates the largest allowed value of the reading.
o
Error
: Indicates the difference between the
Reading
and
Reference
.
o
Error Ratio
: Indicates the ratio of the error to the allowed error band. The closer the
value is to zero, the better the result. If this value is below -100%, then the reading
exceeds its minimum; if above +100%, the reading exceeds the maximum.
o
Pass/Fail
: If the measurements are within tolerance,
Pass
is indicated. If the adjustments
are not within tolerances,
Fail
is indicated.
Save
button: When clicked, the calibration or verification data can be saved to a .csv file.
Clear
button: When clicked, all the data in the results field will be erased.
Содержание Ametek EX1200A Series
Страница 17: ...VTI Instruments Corp 16 EX1200A Series Introduction EX1200A SERIES 1 U FULL RACK MAINFRAME DIMENSIONAL DIAGRAM ...
Страница 18: ...www vtiinstruments com EX1200A Series Introduction 17 EX1200A SERIES 3 U MAINFRAME DIMENSIONAL DIAGRAM ...
Страница 33: ......
Страница 51: ...VTI Instruments Corp 50 EX1200A Series Preparation for Use ...
Страница 64: ...www vtiinstruments com EX1200A Series Preparation for Use 63 ...
Страница 65: ......
Страница 73: ......
Страница 107: ...VTI Instruments Corp 106 EX1200A Triggering ...
Страница 109: ...VTI Instruments Corp 108 EX1200A Triggering FIGURE 6 1 EX1200A SERIES TRIGGER MODEL ...
Страница 143: ...VTI Instruments Corp 142 EX1200A Web Page Operation ...
Страница 145: ...VTI Instruments Corp 144 EX1200A Series Programming ...
Страница 155: ......
Страница 161: ......