www.vtiinstruments.com
EX1200A Triggering
83
X = Resolution/Range
Where Range =100V if userRange=300V, =1A if userRange=3A, = userRange
otherwise
if (X <= 1e-6)
{
numberOfPLC=10;
}
else if (X <= 1e-5)
{
numberOfPLC=1;
}
else if (X <= 1e-4)
{// 4.5 digits
numberOfPLC=0.1;
}
else
{// 3.5 digits
numberOfPLC=0.01;
}
apertureTime=numberOfPLC/PowerLineFrequency;
Note that the EX1200A
DMM’s resolution is related to aperture time and can be modified to set
this property.
NOTE
When the DMM is set to measure ac voltages using
Low/Medium
filter option, the data is
automatically measured with 6½ digit accuracy.
I
NTEGRATION
/A
PERTURE
T
IME
The EX1200A
’s DMM use
s a slope integrating ADC to quantify the analog signal levels into
digital values. The digital bits of resolution are directly proportional to the amount of time used to
integrate the signal.
Integration time is typically specified in number of power line cycles (PLC), except in the case of
frequency and period functions. When integration time is selected as a whole number of PLC, this
helps increase the rejection of common mode and normal mode power line noise. Aperture time is
related to PLC, but is a read-only property and cannot itself be modified. Aperture time is, instead,
controlled by manipulating the resolution. The integration time can be changed by using the
resolution property. This relationship is defined by the algorithm in the
For frequency and period measurements, resolution is set by specifying the aperture time. The
aperture time for the frequency and period functions default to 100 ms. Valid aperture times vary
with the selected resolution: 10 ms for 4½-digits, 100 ms for 5½-digits or 1 s for 6½-digits.
Since the resolution, integration/aperture time, and number of PLC settings are inter-related, some
consideration must be made when configuring these features. The last of these related parameters
to be defined determines the actual resolution of the measurement.
A
UTO
-Z
ERO
The auto-zero function greatly enhances the accuracy of the measurement, especially when the
ambient temperature is constantly changing. When auto-zero is enabled (auto-zero ON), the
multimeter internally disconnects the input signal following each measurement, takes a
“
zero
”
reading, and then a fixed reference reading. These three readings are combined to give a high-
accuracy measurement value. Auto-zero eliminates the negative effects of internal offset and gain
errors that occur with temperature changes and aging. Auto-zero requires two background
measurements so the DUT measurement rate will be reduced. As is true with resolution, however,
performing auto-zero requires two additional measurements to be taken, decreasing the overall
reading rate speed.
Содержание Ametek EX1200A Series
Страница 17: ...VTI Instruments Corp 16 EX1200A Series Introduction EX1200A SERIES 1 U FULL RACK MAINFRAME DIMENSIONAL DIAGRAM ...
Страница 18: ...www vtiinstruments com EX1200A Series Introduction 17 EX1200A SERIES 3 U MAINFRAME DIMENSIONAL DIAGRAM ...
Страница 33: ......
Страница 51: ...VTI Instruments Corp 50 EX1200A Series Preparation for Use ...
Страница 64: ...www vtiinstruments com EX1200A Series Preparation for Use 63 ...
Страница 65: ......
Страница 73: ......
Страница 107: ...VTI Instruments Corp 106 EX1200A Triggering ...
Страница 109: ...VTI Instruments Corp 108 EX1200A Triggering FIGURE 6 1 EX1200A SERIES TRIGGER MODEL ...
Страница 143: ...VTI Instruments Corp 142 EX1200A Web Page Operation ...
Страница 145: ...VTI Instruments Corp 144 EX1200A Series Programming ...
Страница 155: ......
Страница 161: ......