29
Temperature (°C)
EC (µS/cm)
15
119
16
122
17
125
18
127
19
130
20
133
21
136
22
139
23
142
24
145
25
147
26
150
27
153
28
156
29
159
30
162
•
If a different calibration solution is used,
click on the
Set Conduct^ Mod
Calibration Reg 2
command. A new
window will pop up. Input your desired
calibration value and add two times zero
at the end. E.g. if a 300 µS/cm
calibration solution is used delete the
14700 preset value and add 30000,
click OK.
•
The calibration value should be
uncompensated for temperature. If you
don’t know the calibration solution
temperature, you can use the
Read
Temp^ Mod Current Temperature
command to read the calibration
solution temperature, e.g. the 147
µS/cm solution has this EC only at
25°C. If the solution temperature is
20°C the EC value used for calibration
should be 133 µS/cm
•
Click -
Set Conduct^ Mod Calibration
Reg 2
to save the conductivity value of
the calibration solution used.
•
Click
- Set Conduct^ Calibration
Request
-
Single Point
to calibrate the
probe.
•
Click -
Get Conduct^ Mod Calibration
Confirmation
to receive confirmation
•
Click
-
Get
Conduct^
Mod
Conductivity
to get a conductivity
measurement.
•
Click -
Set Conductivity Probe to
Hibernate
.
•
Click
- Reset controller
to save all
changes to the PCB.
•
Disconnect the calibration EC probe
and re-connect the system EC probe on
the PCB. Disconnect the USB cable and
close the cabinet. Jumper should
remain in the open position.