27
•
Connect the external calibration probe.
•
Power on the system.
Conductivity probe is calibrated through
Docklight. Before EC calibration, temperature
has to be calibrated.
Open the program and perform the following by
pressing the
Send
button.
Temperature calibration:
•
Click -
Temp module active
to activate
the probe.
•
Click -
Get Temp^ Mod Calibration.
•
Click –
Clear Temp^ Mod Calibration
to clear any previous calibration
request.
•
Read current room temperature using
an external probe
•
Click on the text of the Set Temp^
Mod Calibration
to set temperature
according to the room temperature
measured in previous step. A pop-up
window will open. By default, the
temperature is set to 25 degrees
Celsius (25000 in Docklight). Replace
this value with the room temperature
measured via the external probe, e.g. if