23
8.4
S
YSTEM
System Branch
Password
To access
System,
you will be asked for a password. The
password is available to authorized dealers.
Setpoints
When advised by Voltea, settings can be adjusted via the LCD.
Navigate to the
Setpoints
tab on the LCD screen.
A cursor shows which digit is being edited. UP increases the value,
and DOWN decreases it. Press ENTER to move onto the next
digit. The edit will only be saved when ENTER is pressed on the
final digit.
Hardware Test
From Hardware Test the Input Read, Relay test and Sensor Test
are accessible.
Input read:
1
Unassigned
2
Bypass
3
Leak sensors
4
Start signal
Input variables can have 2 values, Lo and Hi. Hi when the signal
is received and Lo when there is no signal. E.g.:
•
When the bypass is not active and no leakage while
starting signal is active it should read 1Lo 2Lo 3Lo 4Hi
•
When the bypass is connected it should read 1Lo 2Hi 3Lo
4Hi
•
When there is a leakage it should read 1Lo 2Lo 3Hi 4Hi
Relay test:
Gives access to manual operation for the valves, CIP
pump and relays.
•
MIV: Main inlet valve
•
HFIV: High flow valve
•
POV: Pure outlet valve
•
WOV: Waste outlet valve
•
CIP: CIP pump
•
PVR: Shunt relay
•
NVR: Polarity changing relay
•
PUMP: Power supply
Sensor read:
Gives access to Flow, Conductivity, Water
Temperature and Cabinet temperature live measurements.