28
the probe reads 18
°
C the value in
Docklight should be 18000
Note:
If the probe is not at room
temperature, wait for 5 minutes for it to
reach room temperature before
calibrating. Clicking the Read Temp^
Mod Current Temperature will display
on screen the probe temperature
reading. When this reading is stable
the probe reached room temperature.
•
Click –
Set Temp^ Mod Calibration
Request-Single Point
to calibrate the
temperature.
•
Click –
Get Temp^ Mod Calibration
Confirmation
to confirm calibration
•
Clicking –
Read Temp^ Mod Current
Temperature
will
display
the
temperature reading of the probe.
Conductivity calibration:
•
Click –
Set conductivity Probe Active
to activate the EC probe.
•
Click –
Set Conduct^ Probe Type
•
Click –
Get Conduct^ Probe Type
Response should be “100”.
•
Click
–
Clear
Conduct^
Mod
Calibration Request
to clear previous
calibration request.
•
Click -
Set Conduct^ Mod Calibration
Reg
. Make sure the probe is not in
contact with any liquid or surface.
•
Click -
Set Conduct^ Request-Dry,
to
calibrate for dry.
•
Disconnect the external EC probe from
the PCB and connect the EC probe
mounted on the special manifold.
Fill
the manifold with the calibration
solution. Voltea suggests using a
147µS/cm standard solution.