Vitrolux H
V2
Maintenance
Stand 5/12
Page 44 of 57
5.3.4 Z-axis
The Z axis is lubricated by an opening through the scanner head mounting plate. Please
move the scanner head towards the bottom final position. The cap of the nipple now is visible
through an opening in the mounting plate.
Please remove the cap, lubricate the shaft and replace the plastic cap again. Alternatively
some fat can be applied through the left lip of the seal on the shaft. The shaft is located
about 2 cm left next to the seal.
Figure 3: Access lubrication Z-axis
Plastic cap. Underneath the cap the nipple for lubrication is located
At the end of the lubrication process please plug in the power and switch on again the
system at the main switch.
Please start the control computer and unlock the laser with the green power button again.
Then move all 3 axis by using the "VitroMark" software 2 times over the entire range of
motion to distribute the lubricant.