K465i Operations and Maintenance
P/N 1212831 Rev C
Page 5-168
Copyright
©
2012
Veeco Instruments, Inc. Confidential
All Rights Reserved
MASS FLOW CONTROLLERS AND PRESSURE CONTROLLERS
Note:
Leak-check the MFC and PC by connecting the leak detector downstream of the
device you are checking.
Procedure 5-30: Leak-Checking Mass Flow Controllers
Note:
Always leak check mass flow controllers the leak detector hooked up to the
downstream side.
1. Give the MFC a full setpoint to open it.
2. Spray helium around the body of the MFC.
3. If a leak is detected, refer to the above VCR section for mitigating the leak. If checking the VCR fitting
does not mitigate the leak, check the o-
ring and replace as necessary (on elastomer sealed MFC’s).
Procedure 5-31: Leak-Checking a Pressure Controller
Note:
Always leak check pressure controllers from both sides; that is, with the leak
detector hooked up to each side, individually.
.
1. Spray helium around the body of the pressure controller.
2. If a leak is detected, refer to the above VCR section for mitigating the leak. If checking the VCR fitting
does not mitigate the leak, check the o-ring and replace as necessary (on elastomer sealed pressure
controllers).