15. Using a wipe and ESD (the approved) vacuum cleaner, loosen the particles from all stainless steel
surfaces and vacuum them (without making contact with stainless steel).
16. Loosen the belly band below the second stage until it can be removed from the filter body of the
particle filter housing. Lift the housing piece.
Warning! Watch for fire or flare ups.
Figure 5-135. Lower (1st) Stage
), lift the filter stage out of the housing and place the assembly
onto a table or maintenance area.
18. Using a wipe and ESD (the approved) vacuum cleaner, loosen the particles from all stainless steel
surfaces and vacuum them (without making contact with stainless steel).
19. Loosen the belly band below the second stage until it can be removed from the filter body of the
particle filter housing. Lift the housing piece.
Warning! Watch for fire or flare ups.
The first stage is now fully exposed and ready to clean.
20. Using a wipe and ESD (the approved) vacuum cleaner, loosen the particles from all stainless steel
surfaces and vacuum them (without making contact to stainless steel). IPA may be used to remove
hardened alkyl deposits.
Particle Filter Assembly
21. Prepare the maintenance kit.
22. Place the filter inside the second stage housing, concentric to the raised lip.