4. Imaging System Overview
36
Green Smart User Manual
4.6.3 Emergency Stop Switch
During operation, the following emergency situations may occur:
X-ray emission even after the
Exposure Switch
has been released
Physical injury to the patient or damage to the equipment
Other emergency situations
If a problem occurs during image acquisition, press the red
Emergency Stop Switch
to
immediately stop the moving parts and cut off all power to the equipment’s electrical
components. To restart the equipment, turn the
Emergency Stop Switch
clockwise until it
pops up.
The
Emergency Stop Switch
is located at the bottom of the Vertical Frame.
4.6.4 Exposure Switch
The
Exposure Switch
allows the operator to control image acquisition from outside of the
X-ray room.
Press and hold the
Exposure Switch
down until acquisition is completed. Premature
release of the
Exposure Switch
will abort image acquisition.
Pressing the
Exposure Switch
activates the LED indicator to turn yellow. This color
indicates that the X-ray is being emitted.
The
Exposure Switch
is detachable. Ensure the
Exposure
Switch
cable is not detached from the unit accidentally during the
operation.
Keep vocal/visual contact with the patient during exposure. If any
problem occurs during exposure, release the
Exposure Switch
immediately.
Содержание GREEN SMART
Страница 1: ...User Manual Version 1 23 English...
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 52: ...5 Imaging Software Overview 44 Green Smart User Manual Left blank intentionally...
Страница 90: ...7 Acquiring PANO Images 82 Green Smart User Manual Left blank intentionally...
Страница 110: ...8 Acquiring CEPH Images Optional 102 Green Smart User Manual Left blank intentionally...
Страница 126: ...9 Acquiring CBCT Images 118 Green Smart User Manual Left blank intentionally...
Страница 134: ...10 Acquiring 3D MODEL Scan Images 126 Green Smart User Manual Left blank intentionally...
Страница 142: ...12 Cleaning and Maintenance 134 Green Smart User Manual Left blank intentionally...
Страница 144: ...13 Disposing of the Equipment 136 Green Smart User Manual Left blank intentionally...
Страница 149: ...14 Technical Specifications Green Smart User Manual 141 ENGLISH Maximum Rating Charts Emission Filament Characteristics...
Страница 154: ...14 Technical Specifications 146 Green Smart User Manual Left blank intentionally...
Страница 185: ...15 Appendices Green Smart User Manual 177 ENGLISH...
Страница 198: ...15 Appendices 190 Green Smart User Manual STL Stereo Lithography TMJ Temporomandibular Joint UHD Ultra High Definition...
Страница 200: ......