5. Imaging Software Overview
42
Green Smart User Manual
9
CONFIRM /
READY
button
Applies the selected settings and moves to the next step.
(Exposure parameter setting and patient positioning > Ready
for exposure)
When you click
CONFIRM
button, estimated
DAP (Dose Area Product), scan time and
exposure time would be displayed DAP, Scan
Time and Exposure Time Display window.
Activated when you click
CONFIRM
button after the patient
positioning is completed. Click the button when all aspects of
preparation are completed for image acquisition.
10
Imaging
parameters
configuration
panel
Selects the imaging parameters for each mode: PANO, CEPH
(Optional), CBCT and 3D MODEL Scan
11
Modality
Selection
button
Returns to Modality Selection (PANO, CEPH (Optional), CBCT
and MODEL) screen.
12
Rotation Test
button
Switches to the Rotation Test mode to check if any part of the
patient’s body reaches the surface of the equipment before the
actual exposure.
To change to the Rotation test mode,
1. Align the patient to the equipment. (For the details, refer
to the “Positioning the Patient” section of each modality
chapter.)
2. Select a modality.
3. Click
CONFIRM
button.
4. Click
Rotation Test
button. Then, the button is changed
to OFF state and icon.
To start the rotation test, press
BEAM ON/OFF
button on
the Control Panel.
To finish the test mode, click
Rotation Test
button or
READY
button.
This function is applied to PANO and CBCT
modalities only.
Содержание GREEN SMART
Страница 1: ...User Manual Version 1 23 English...
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 52: ...5 Imaging Software Overview 44 Green Smart User Manual Left blank intentionally...
Страница 90: ...7 Acquiring PANO Images 82 Green Smart User Manual Left blank intentionally...
Страница 110: ...8 Acquiring CEPH Images Optional 102 Green Smart User Manual Left blank intentionally...
Страница 126: ...9 Acquiring CBCT Images 118 Green Smart User Manual Left blank intentionally...
Страница 134: ...10 Acquiring 3D MODEL Scan Images 126 Green Smart User Manual Left blank intentionally...
Страница 142: ...12 Cleaning and Maintenance 134 Green Smart User Manual Left blank intentionally...
Страница 144: ...13 Disposing of the Equipment 136 Green Smart User Manual Left blank intentionally...
Страница 149: ...14 Technical Specifications Green Smart User Manual 141 ENGLISH Maximum Rating Charts Emission Filament Characteristics...
Страница 154: ...14 Technical Specifications 146 Green Smart User Manual Left blank intentionally...
Страница 185: ...15 Appendices Green Smart User Manual 177 ENGLISH...
Страница 198: ...15 Appendices 190 Green Smart User Manual STL Stereo Lithography TMJ Temporomandibular Joint UHD Ultra High Definition...
Страница 200: ......