8. Acquiring CEPH Images (Optional)
100
Green Smart User Manual
8.4
X-ray Exposure
If an emergency occurs during image acquisition, release the
Exposure Switch
to cease X-ray emission.
The operator shall observe the X-ray safety regulations applicable
to his/her area at all times during the operation of this equipment.
The operator must keep vocal/visual contact with the patient at all
times during the image acquisition process.
Do not operate the PC during exposure. Doing so may cause the
system to malfunction.
Let the patient close the eyes during the operation.
To acquire optimized images, instruct the patient to hold his/her
breath and not to swallow. Also, don’t let the patient move until the
Temple Supports are open.
1.
Get out of the X-ray room and close the door.
2.
Press and hold down the
Exposure Switch
until image acquisition is completed.
The image appears on the screen.
During X-ray exposure, the status appears as follows.
The LED light of the
Exposure Switch
turns yellow.
The LED light on the top of the equipment turns yellow.
An alert sound comes out to indicate that X-ray emission is
currently underway.
On Console Software, the radiation mark turns yellow and “X-RAY”
changes to “X-RAY ON.”
3.
Release the
Exposure Switch
when “Image capturing is completed” message appears on
the screen.
Содержание GREEN SMART
Страница 1: ...User Manual Version 1 23 English...
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 52: ...5 Imaging Software Overview 44 Green Smart User Manual Left blank intentionally...
Страница 90: ...7 Acquiring PANO Images 82 Green Smart User Manual Left blank intentionally...
Страница 110: ...8 Acquiring CEPH Images Optional 102 Green Smart User Manual Left blank intentionally...
Страница 126: ...9 Acquiring CBCT Images 118 Green Smart User Manual Left blank intentionally...
Страница 134: ...10 Acquiring 3D MODEL Scan Images 126 Green Smart User Manual Left blank intentionally...
Страница 142: ...12 Cleaning and Maintenance 134 Green Smart User Manual Left blank intentionally...
Страница 144: ...13 Disposing of the Equipment 136 Green Smart User Manual Left blank intentionally...
Страница 149: ...14 Technical Specifications Green Smart User Manual 141 ENGLISH Maximum Rating Charts Emission Filament Characteristics...
Страница 154: ...14 Technical Specifications 146 Green Smart User Manual Left blank intentionally...
Страница 185: ...15 Appendices Green Smart User Manual 177 ENGLISH...
Страница 198: ...15 Appendices 190 Green Smart User Manual STL Stereo Lithography TMJ Temporomandibular Joint UHD Ultra High Definition...
Страница 200: ......