11. Troubleshooting
Green Smart User Manual 129
EN
G
LISH
Error Code
Description
equal to 200 V upon X-ray emission. (In case of less than or equal
to 200 V, replace the power source.)
- After replacing the inverter board, check for error after X-ray
emission. If the same error occurs, replace the monk tank.
H013
Error occurs when the kV feedback is greater than or equal to the
set value by more than or equal to 20 kV during X-ray emission
(kV feedback ≥ set value + 20). The number "4" appears on the
inverter board.
- After emitting X-rays again, check whether the same error
occurs.
- Check the input power (DC 300 V) of the inverter board.
- Check that the input power (DC 300 V) is changed to less than
200 V upon X-ray emission. (In case of less than 200 V, replace
the power source.)
- After replacing the inverter board, check for error after X-ray
emission. If the same error occurs, replace the monk tank.
H014
Error occurs when the mA feedback is less than or equal to the
50 % of the set value during X
-ray emission (mA feedb
ack ≤ set
value x 0.5). The number "6" appears on the inverter board.
- After emitting X-rays again, check whether the same error
occurs.
- Check the input power (DC 300 V) of the inverter board.
- After replacing the inverter board, check for error after X-ray
emission. If the same error occurs, replace the monk tank.
H015
Error occurs when the mA feedback is greater than or equal to the
150 % of the set value during X
-ray emission
(mA feedback ≥ set
value x 1.5). The number "6" appears on the inverter board.
- After emitting X-rays again, check whether the same error
occurs.
- Check the input power (DC 300 V) of the inverter board.
- After replacing the inverter board, check for error after X-ray
emission. If the same error occurs, replace the monk tank.
H020
Error occurs during movement from the P-axis motor's origin
position.
- Check the connection lines for the motor and origin sensor. If any
problem, replace the motor or sensor.
H021
Error occurs during movement from the rotator-axis motor's origin
position.
- Check the connection lines for the motor and origin sensor. If any
problem, replace the motor or sensor.
H024
Error occurs during movement from the CT horizontal laser
up/down-axis motor's origin position.
- Check the connection lines for the motor and origin sensor. If any
problem, replace the motor or sensor.
H027
Error occurs during movement from the Cephalo sensor motor's
Содержание GREEN SMART
Страница 1: ...User Manual Version 1 23 English...
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 52: ...5 Imaging Software Overview 44 Green Smart User Manual Left blank intentionally...
Страница 90: ...7 Acquiring PANO Images 82 Green Smart User Manual Left blank intentionally...
Страница 110: ...8 Acquiring CEPH Images Optional 102 Green Smart User Manual Left blank intentionally...
Страница 126: ...9 Acquiring CBCT Images 118 Green Smart User Manual Left blank intentionally...
Страница 134: ...10 Acquiring 3D MODEL Scan Images 126 Green Smart User Manual Left blank intentionally...
Страница 142: ...12 Cleaning and Maintenance 134 Green Smart User Manual Left blank intentionally...
Страница 144: ...13 Disposing of the Equipment 136 Green Smart User Manual Left blank intentionally...
Страница 149: ...14 Technical Specifications Green Smart User Manual 141 ENGLISH Maximum Rating Charts Emission Filament Characteristics...
Страница 154: ...14 Technical Specifications 146 Green Smart User Manual Left blank intentionally...
Страница 185: ...15 Appendices Green Smart User Manual 177 ENGLISH...
Страница 198: ...15 Appendices 190 Green Smart User Manual STL Stereo Lithography TMJ Temporomandibular Joint UHD Ultra High Definition...
Страница 200: ......