TECHNICAL INFORMATION
84/94
CN929-7015M002
E-Net Dual Controller Error Status
Value
Error type
Error reference
High Voltage errors
1
31h
High Voltage off due to front panel interlock
activation
Panel Interlock
2
32h
High Voltage off due to Remote I/O interlock
activation
Remote Interlock
3
33h
High Voltage off due to Cable HV interlock
activation
Cable Interlock
4
34h
Dual fault
HV not found
5
35h
High Voltage off due to a general DSP
determined fault
HV Fault
6
36h
High Voltage off due to an HV module
overtemperature determined by the DSP
HV Overtemperature
7
37h
Remote I/O card not present or faulty
R.I/O not found
8
38h
Remote I/O card present, but faulty
R.I/O fault
9
39h
High Voltage off due to the protect function
activation
Protect
10
31h30h
High Voltage off due to shortcircuit protection
activation
Short Circuit
11
31h31h
High Voltage off due to an HV module
overvoltage or overcurrent determined by the
DSP
Over Volt/Curr
12
31h32h
High Voltage off due to the zero measurement
protection activation
Zero Meas
System errors
1
31h
RAM failure: RAM diagnostics error
Software Error
2
32h
config register: incorrect value in the uC
68HC11 configuration register
Software Error
3
33h
test mode: invalid uC 68HC11 operating mode
Software Error
4
34h
copyright: violation of the signature in the ROM
or the ROM was corrupted
Software Error
5
35h
eeprom fault: checksum or non-volatile memory
write errors. Factory defaults are automatically
loaded
Software Error
6
36h
version number: incompatible uC and Dsp
versions
Software Error
7
37h
hv dsp not found: the Dsp does not respond
during the uC initialization phase
Software Error
8
38h
dsp fault: the Dsp does not respond during
normal operation
Software Error
9
39h
invalid option: option card not configured
correctly
Software Error
10
31h30h
unknow option: generic execution error
Software Error