INSTRUCTIONS FOR USE
10/94
CN929-7015M002
To attach the controller to the pump use the specific
cable supplied as option.
Refer to the appendix entitled “Technical Information”
for detailed information on these and other connections.
Ground Connections
NOTE
The cable interlock closes on pump ground. The high
voltage is disabled if the connection is interrupted.
Close the interlock using the appropriate counter-
connector when using a cable which is not equipped
with the interlock connector.
USAGE
This section provides you with the major operating
procedures. For more information and for the
procedures that concern connections or specific
options, refer to the section “USAGE” in the “Technical
Information” appendix. Before using the controller,
perform all electrical and pneumatic connections and
refer to the manual of the connected pump.
!
To avoid injury to persons or damage to the equipment,
if the pump needs to be placed on a table make sure
that the table is stable. Never activate the pump if the
input flange is not connected to the system or if it is not
closed with the blank-off flange.
Controls and Indicators on the E-Net Dual
Controller Front Panel
The following figure shows the E-Net Dual Controller
front panel. The following table gives the definition and
function of the controls and indicators.
Front Control Panel
Legend:
1.
Back-lit Liquid Crystal Display, 16 characters x 2 rows
2.
Green LEDs, indicating the selected operating mode:
- Local LED on when the LOCAL operating mode
is selected
- Serial LED on when the SERIAL operating mode
is selected
- Remote LED on when the REMOTE I/O operating
mode is selected
DANGER!