TECHNICAL INFORMATION
76/94
CN929-7015M002
LIST OF COMMANDS
The commands that have been implemented are described in the following tables:
General Commands
Command
Description
Mode
Channels
Format
Possible Values
Local/remote
Local or serial
operating mode
Read/Write
No channel
Status
30h local
31h Remote I/O
32h serial
HV on/off
On/off
High Voltage
Read/Write
HV1, HV2
Status
See HV on/off command
in the codes table
Unit of measure
Unit of measure
on E-Net Dual
Controller
Read/Write
No channel
Status
30h Torr
31h mBar
32h Pascal
µC firmware
version
µC (micro central)
firmware version
Read
No channel
String
18 characters
e.g. “VPo 1 0 24/04/98”
DSP firmware
version
Versione firmware
DSP (micro HV)
Read
No channel
String
16 characters
e.g. “VPd 1 0 24/04/98”
Device number
Pump number set
on the specified
channel
Read/Write
HV1, HV2
Gauge1, Gauge2
Serial
Status
See device table
Device type
Type of device set
on the specified
channel
Read
Channel no.
HV1, HV2
Gauge1, Gauge2
Serial
String
See device table
V meas
Voltage
measurement
Read
HV1, HV2
Integer
(V)
“00000” if powered off
I meas
Current
measurement
Read
HV1, HV2
Exp. (A)
“0 0E+00” if powered off
Pr meas
Pressure
measurement
Read
HV1, HV2
Gauge1, Gauge2
Exp.
(Torr)
“0 0E+00” if powered off
E-Net Dual
Controller
Error Status
Reads the error
status on the
selected channel
Read
HV1, HV2
Gauge1, Gauge2
Serial
no channel
Integer
See Error Status Table
Serial reset
Resets the E-Net
Dual Controller
Write1
No channel
Status
30h not ok,
no reset
31h ok, reset performed
Remote error
Error returned by
the E-Net Dual
Controller due to
an invalid data
exchange
Reply
No channel
Integer
See the Protocol error
table
Interlock
Status
Reads the status
of the interlocks
Read
No channel
BitField
See the Interlock table
1
E-Net Dual Controller doesn't reply to this command.