TECHNICAL INFORMATION
56/94
CN929-7015M002
CONTROLLER SPECIFICATIONS
Input:
Voltage
Frequency
Power
100 to 120 Vac / 200 to 240 Vac
50 – 60 Hz
600 VA maximum
Output for pump:
Voltage
Short circuit current
Power
±
7,000 Vdc +2 –10% maximum
programmable via software between 3,000 V and
7,000 V at 100 V steps
400 mA
400 W maximum (programmable via software between 100
W and 400 W at 10 W steps)
Output current
400 mA
±
10% (programmable via software between 100
mA and 400 mA at 10 mA steps)
Operating temperature
0 °C to +35 °C
Storage temperature
-20 °C to +70 °C
Fuse:
Line voltage
Two 4 A(slow blow) for 200 or 240 input Voltage
Two 8A (slow blow) for 100 or 120 input Voltage
Voltage measurement
Resolution
≤
50 V
Accuracy
±
2% of Vmax
Current measurement
Measurement range: 10 nA to 400 mA
Resolution: 10 nA
Accuracy: I >10
µ
A
±
2%, I < 10
µ
A
±
10%
Pressure Analog REC. OUT
0 to +5 Vdc log. See Rec Out diagram. Minimum recorder
input impedence 1 MOhm
Voltage Analog REC. OUT
0 to +7 Vdc line. See Rec Out diagram. Minimum recorder
input impedence 1 MOhm
Measurement updates
On the display: t
r
≤
2 s
On the recorder: t
r
≤
0.1 s
In compliance with norms
EN 55011 class A group 1
EN 61000-3-2 (Harmonic Current Emission)
EN 61000-3-3 (Flicker)
EN 61000-4-2 (ESD)
EN 61000-4.4 (Burst)
EN 61000-4.3 (Radiated RF immunity)
EN 61010-1 (Safety)
Installation category
II
Pollution degree
2
Interconnecting cables
Power cord (3-wire, 3-meter long) with European or US
plug
Weight (both models)
With 1 H.V.10 Kg (22.4 lbs)
With 2 H.V. 11.5 Kg (25.8 lbs)