TECHNICAL INFORMATION
72/94
CN929-7015M002
Set-point 1 or Set-point 2 selection
(valid for all pumps)
H V 1
S e t P t 1 ( T o r
r
)
→
1 .
0
E X P
- 6
*
At this point it is possible to use the CHANGE and
ENTER buttons to change and confirm the values for
the mantissa and exponent.
Pressing the ENTER button to confirm the selection
made returns you to the menu of the voltages on the
two channels.
Messages Associated to the Menu Button
Related to the Serial Channel
The serial menu described below can be accessed if
SERIAL was selected during the channel selection
phase.
As for the previous menu, also in this case you can use
the CHANGE button to change the parameters and the
ENTER button to confirm them.
Baud Rate Selection
S C 1
B A U D
R A T E
9 6 0 0
It is possible to select between 9600, 4800, 2400, and
1200
Address Selection
S C 1
A D D R E S S
N O D E
1
*
Parity Selection
S C 1
P A R I T Y
N O N E
*
The following values can be chosen:
NONE
ODD
EVEN
SERIAL INTERFACE
General
After having described the connector and the pinouts in
the different configurations, this section provides
information on E-Net Dual Controller’s serial interface.
Hardware Signals Management
The RX, TX and GND signals are present on the data
exchange line.
There is no flow control implemented through RTS/CTS
and the DTR and DSR signals are not handled.
Transmission Format and Timeout
The serial data transmission parameters are listed
below:
Speed
1200-2400-4800-9600 baud
1
Start bit
1
Stop bit
1
Data bit
8
Parity
None, Odd, Even
1
All the baud rates indicated are supported; the default
rate is 9600.
Data exchange is performed in the Half-Duplex mode.
To guarantee a synchronized data flow and to prevent
conflicts, the timings indicated in the table below must
be respected: