![Valco baby Vici TCD-3 Скачать руководство пользователя страница 40](http://html.mh-extra.com/html/valco-baby/vici-tcd-3/vici-tcd-3_instruction-manual_849548040.webp)
36
Appendix C: Heater and Filament Connector
Signal Descriptions and Connection Matrices
Heater Connector
5
4
3
2
1
n.c. Filament B- Filament B- Filament B+ Filament B+
B
n.c. Filament A- Filament A- Filament A+ Filament A+
A
4
3
2
1
HTR- HTR+
RTD- RTD+
B
HTR- CHSGND RTD- RTD+
A
pin
/signal
B4
/HTR-
B3
/HTR+
B2
/RTD-
B1
/RTD+
A4
/HTR-
A3
/CHSGND
A2
/RTD-
A1
/TRD+
B4
/HTR-
–
40
Ω
open
open
SHORT
open
open
open
B3
/HTR+
40
Ω
–
open
open
40
Ω
open
open
open
B2
/RTD-
open
open
–
1100
Ω
open
open
SHORT
1100
Ω
B1
/RTD+
open
open
1100
Ω
–
open
open
1100
Ω
SHORT
A4
/HTR-
SHORT
40
Ω
open
open
–
open
open
open
A3
/CHSGND
open
open
open
open
open
–
open
open
A2
/RTD-
open
open
SHORT
1100
Ω
open
open
–
1100
Ω
A1
/RTD+
open
open
1100
Ω
SHORT
open
open
1100
Ω
–
pin
/signal
B5
/n.c.
B4
/FIL B-
B3
/FIL B-
B2
/FIL B+
B1
/FIL B+
A5
/n.c.
A4
/FIL A-
A3
/FIL A-
A2
/FIL A+
A1
/FIL A+
B5
/n.c.
–
open
open
open
open
open
open
open
open
open
B4
/FIL B-
open
–
SHORT
25
Ω
25
Ω
open
open
open
open
open
B3
/FIL B-
open
SHORT
–
25
Ω
25
Ω
open
open
open
open
open
B2
/FIL B+
open
25
Ω
25
Ω
–
SHORT
open
open
open
open
open
B1
/FIL B+
open
25
Ω
25
Ω
SHORT
–
open
open
open
open
open
A5
/n.c.
open
open
open
open
open
–
open
open
open
open
A4
/FIL A-
open
open
open
open
open
open
–
SHORT
25
Ω
25
Ω
A3
/FIL A-
open
open
open
open
open
open
SHORT
–
25
Ω
25
Ω
A2
/FIL A+
open
open
open
open
open
open
25
Ω
25
Ω
–
SHORT
A1
/FIL A+
open
open
open
open
open
open
25
Ω
25
Ω
SHORT
–
Filament Connector
Signal definitions
Signal definitions
Connection matrix
Connection matrix
NOTES
Connections marked as "open" should have no connectivity, and a measured resistance over 10 M
Ω
.
SHORT connections are measurement loops, and should have a measured resistance under 1
Ω
.
25
Ω
Measurements across the filaments, at 25°C. Resistance will increase with temperature, to as much as 75
Ω
at
temperatures exceeding 300°C.
40
Ω
Measurements across the heater element, for a 24V system at 25°C.
1100
Ω
Measurements across the 1000
Ω
platinum RTD sensor. The 1100
Ω
value is nominal for the RTD at 25°C.
Содержание Vici TCD-3
Страница 2: ...This page intentionally left blank for printing purposes ...
Страница 42: ...38 Detector Performance Log ...