Figure 8:
TCD-3 cable connections
Installation
Flow Rate Settings
Since the TCD-3 is a concentration-dependent detector, the lower the flow rate
through the detector, the higher the sensitivity. Column diameter will determine
the optimum column flow rate.
The microvolume TCD-3 is designed for the lower flow rates typical for capillary
columns, and achieves best sensitivity at rates below 10 mL/min. Since the fila-
ments are maintained at constant temperature, the detector can be operated at
extremely low flow rates (less than 0.5 mL/min) without damage to the filaments.
1. Measure the column flow at the detector’s Channel A out. The optimum flow
rate is in the range of 4 – 20 mL/min, with the actual rate dependent on the
type of column used.
2. Measure the reference gas flow at Channel B out. It should be as close as
possible to the GC column flow. Use a fixed flow restrictor or a good quality
flow controller to match the carrier gas and reference gas flows.
3. Once the flows have been established, connect the 48VDC power supply to the
TCD controller unit, plug it into an appropriate source, and turn the power on
via the rear panel switch.