![Valco baby Vici TCD-3 Скачать руководство пользователя страница 27](http://html.mh-extra.com/html/valco-baby/vici-tcd-3/vici-tcd-3_instruction-manual_849548027.webp)
23
4. Graph
icon
Clicking the graph opens the Graph screen
(page 26)
.
5. Gear
icon
Clicking the gear on the Control or Graph screen returns the GUI to this
Configuration screen. On the Configuration screen, it has no functionality.
6. Question Mark
icon
Clicking the question mark opens a help section.
7. X
icon
Click to close the TCD-3 GUI program.
8. Detector Status
info box
This box is empty until communication has been established. Then it will
display one of the messages below:
Detector not found
The detector unit was not sensed by the TCD-3
controller. Check that the detector unit is connected
to the controller, and that both cables are fully seated
in their sockets.
Temperature not yet stable
The detector has not yet reached full temperature
stability. Note that the minimum time to stability is
30 minutes.
Detector stable
The detector has reached thermal stability, and is
ready for analysis.
Table 6:
Possible messages in the Detector Status info box
Figure 13
: Configuration screen of the TCD-3 GUI
TCD Operation with the Graphical User Interface