Installing and Purging a Purifier
1. If the pressure regulator has a 1/8" male cone-type outlet port, install a Valco
1/8" external to 1/16" internal reducer (EZR21); if it has a 1/4" male cone-type
outlet port, install a Valco 1/4" external to 1/16" internal reducer (EZR41). For
other regulator outlet fittings, a wide variety of Valco adapters are available.
2. While keeping the outlet tube capped, remove the cap from the inlet tube
of the Valco helium purifier and insert the tube fitting into the 1/16" reducer
port. Use a 1/4" wrench to turn the nut one-quarter turn past the point where
the ferrule first starts to grab the tubing. Do not remove the fitting. When
made up properly, it should be leak-tight.
3. Turn the output pressure regulating knob clockwise until the gauge registers
345 KPA (50 psi).
4. Allow five minutes for equilibration, then turn the regulating knob all the way
counterclockwise.
5. Observe the needle of the output pressure gauge for 15 minutes. There will
be a slight initial drop, but if it doesn’t move after that, consider that all the
connections are tight.
6. If necessary, use an electronic leak detector to locate any leaks. If a leak
detector is not available, tighten all the fittings (including the output
pressure gauge), and repressurize the system for another test.
7. Uncap the outlet tube of the purifier and purge the system for 15 to 30 minutes
at 60 to 80 mL/min to eliminate air from the purifier getter material.
Column Connection
To prevent detector contamination, it is strongly
recommended that the column be disconnected
from the detector during column bake-out procedures.
Referenced mode
In the referenced mode, the column is connected to the Channel A inlet, and the
carrier gas stream is split and used as a reference in channel B. The gas flow rate
from both channels must be the same (flow balanced). This mode of operation
provides the best baseline stability and the least background noise.
Unreferenced mode
In the unreferenced mode, Channels A and B are used independently. One column
connects to the Channel A inlet, and the other to Channel B. The carrier gas must
be the same for both channels. It is also possible to use only one channel with
the other capped off, as long as the filament temperature control of the unused
channel is set to minimum.
The unreferenced mode has certain limitations. There is a possibility of detector
cross-talk (interference) if compound concentrations exceed 1%. Also, baseline
drift and background noise are greater compared with the referenced mode.
EZR21
Installation