![Valco baby Vici TCD-3 Скачать руководство пользователя страница 19](http://html.mh-extra.com/html/valco-baby/vici-tcd-3/vici-tcd-3_instruction-manual_849548019.webp)
15
Cable Connections
Before connecting the any detector cables to the control
module, make sure that the control module power cord
is unplugged and the TCD3 controller power switch is
off. Hot-plugging system elements is not recommended.
NOTE: The HEATER and FILAMENTS connectors carry
live voltages, and should only be used with equipment
having no accessible live parts.
Heater and Filament Connections
The heater and filament connections from the detector to the rear panel of the
controller are made via two cables. (
Figure 8
) Cable functions are clearly marked
on the detector, as are the connectors are the controller.
After making sure that the controller power cord is not plugged in, connect the
cables to the appropriate connectors on the control module. The cable connectors
should be inserted into the controller sockets until they latch securely into place.
Pin assigments are indicated below.
Data Output Cable(s)
Connect the data output cable(s). The A / B / A-B outputs are typically used with
an integrator or PC-based data acquisition system. Polarities and full scale range
are indicated in
Figure 8
. Set the recorder (DAQ) for the same voltage scale.
Serial Data Extension Cable
The TCD-3 connects to a PC/laptop via a RS-232C connection. The provided DE-9F
port on the rear panel of the TCD3-C unit uses industry standard (EIA-232-C)
pin-outs for a DCE device. Connect this port to a standard PC serial port using a
standard RS-232C serial data extension cable (male to female).
Digital I/O Cable (optional)
In addition to the serial communication port described above, the TCD-3 controller
also provides digital signals for monitoring and control of the system.
Refer to
“Digital I/O Control” on page 30.
Power Cable
Plug the 4' cable from the 48 VDC power supply into the connector labeled 48VDC
on the rear panel of the controller, but do not plug the power supply into the 110
or 230 VAC source.